Pipelle mark II

Cat #

111020100

규격

box of 25 units

원산지

France

제조사

CCD

 

설명

The Pipelle® Mark II includes:

- A flexible, transparent polypropylene sheath, 3.10 mm in external diameter, 2.6 mm in internal diameter and 23.5 cm long, with a lateral orifice 2.1 mm in diameter in its distal portion and four guide-marks 4, 7, 8 and 10 cm from its extremity. The sheath’s distal extremity is ground on 2 cm.
- An internal EVA piston, which slides up and down when pushed by a flexible acetal resin shaft.

Class I medical device, complying with Directive 93/42/EEC.
Single use.
Latex free.
Sterilised with ethylene oxide

적용

 - The Pipelle® Mark II can be shaped before taking it out of its sterile packaging. Its slight elastic resistance helps the device retain a convexity to fit a uterine anteflexion or retroflexion.
- Disinfect the cervix thoroughly.
- In most cases, Pozzi forceps are not necessary.
- In post-menopausal women, a local anesthesia with xylocaine helps clear a stenosed cervix.
- Slide the Pipelle® Mark II gently through the cervix and up to the uterine fundus. The 4 guide-mark indicates the beginning of the uterine cavity. The 7 guide-mark can be reached in most cases, unless there is a cervical obstacle.
- Draw back the plunger fully to create a negative pressure.
- Sweep the uterine fundus slowly several times up to the internal orifice of the cervix, using regular to-and-fro movements while rotating the sampler to include the whole uterine cavity in the specimen.
- Continue until fragments of uterine mucosa appear within the sheath, which generally takes 30 seconds.
- Remove the Pipelle® Mark II fully and recover the cytology material first, then the histology material, as follows:

Cytology:

- Liquid-based cytology: cut the rough distal tip of the Pipelle® Mark II (approximately 2 cm) and place in a vial containing a solution for liquid-based cytology.

- Smear-based cytology: smear the specimen collected on the rough section onto a slide and fix.

Histology:

- Push the plunger to expel the content of the sheath in a solution for histology.

 

                                                                                Home 으로

 

사업자 등록번호 :204-81-39770  | 사업장주소 : 서울 서초구 서초동 1659-20 세영빌딩

(주)세영종합무역상사 대표이사 : 정정길 | 개인정보관리책임자 : 이승길 | 문의 02)3474-3074 | 팩스 : 02)3474-3240

대구지사 : 대구광역시 동구 신천 동 123-1 농공빌딩 Tel : 053)745-4074  Fax :053)755-4287

seyoung@seyoungls.com    (주)세영종합무역상사  Tel: 02)3474-3074   Fax: 02)3474-3240