1) Sample volume 0.2 to 0.25ml
2) Test time
-low range:approximately 2minutes
-high range:aproximately 3minutes
3) Sample capacity : single sample
4) Units : mOsm/kgH2O
5) Resolution : mOsm/kgH2O
6) Range
-low range 0 to 2000 mOsm/kgH2O
-high range 1400 to 4000 mOsm/kg H2O
7) Communitations : on-board printer, DTE RS-232, and optional barcode scanner
8) Linearity : less than ±0.5%(straight-line method)
9) Repeatability
-0-400 mOsm/kg H2O:±2mOsm (1S.D.)
-400-4000 mOsm/kg H2O:±0.5mOsm (1S.D.)
10) Drift : less than 1 mOsm/kg H2O per month
11) Temperature effects : less than 1 mOsm/kg H2O per 5¡É(9£Æ) ambient temperature change
12) Temperature : 18 to 35¡É(64 to 95£Æ)
13) Relative humidity : 5 to 80% R.H., non-condensing
14) Storage temperature : -40 to +45¡É(-40 to +113£Æ)
15) Electrical :
-voltage : 100 to 130 VAC(1.2A), (50/60Hz) or 200 to 250 VAC(0.6A), (50/60Hz)
-power consumption : 150Watts
16) Dimensions : 16"Hx13"Wx18"D(41cmx32cmx46cm)
|